Obsolete and Discontinued Commands
35
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1305
:TRIGger:TV:TVMode
Command Syntax
:TRIGger:TV:TVMode <mode>
<mode> ::= {FIEld1 | FIEld2 | AFIelds | ALINes | LINE | VERTical
| LFIeld1 | LFIeld2 | LALTernate | LVERtical}
The :TRIGger:TV:MODE command selects the TV trigger mode and field. The
LVERtical parameter is only available when :TRIGger:TV:STANdard is GENeric. The
LALTernate parameter is not available when :TRIGger:TV:STANdard is GENeric (see
).
Old forms for <mode> are accepted:
Query Syntax
:TRIGger:TV:TVMode?
The :TRIGger:TV:TVMode? query returns the TV trigger mode.
Return Format
<value><NL>
<value> ::= {FIE1 | FIE2 | AFI | ALIN | LINE | VERT | LFI1 | LFI2
| LALT | LVER}
<mode>
Old Forms Accepted
FIEld1
F1
FIEld2
F2
AFIeld
ALLFields, ALLFLDS
ALINes
ALLLines
LFIeld1
LINEF1, LINEFIELD1
LFIeld2
LINEF2, LINEFIELD2
LALTernate
LINEAlt
LVERtical
LINEVert
NOTE
The :TRIGger:TV:TVMode command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :TRIGger:TV:MODE command (see
) instead.