:FUNCtion<m> Commands
17
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
411
:FUNCtion<m>:TRENd:NMEasurement
Command Syntax
:FUNCtion<m>:TRENd:NMEasurement MEAS<n>
<n> ::= # of installed measurement, from 1 to 8
<m> ::= 1 to (# math functions) in NR1 format
The :FUNCtion<m>:TRENd:NMEasurement command selects the measurement
whose trend is shown in the math waveform.
There are 8 locations (or slots) that installed measurements can occupy. The
MEAS<n> parameter specifies the location of the measurement whose trend you
want to analyze.
You can view the trend math function waveform for these installed measurements:
• :MEASure:VAVerage, <interval> ::= CYCLe
• :MEASure:VRMS, <type> ::= AC (AC RMS)
• :MEASure:VRATio, <interval> ::= CYCLe
• :MEASure:PERiod
• :MEASure:FREQuency
• :MEASure:PWIDth
• :MEASure:NWIDth
• :MEASure:DUTYcycle
• :MEASure:NDUTy
• :MEASure:RISetime
• :MEASure:FALLtime
Query Syntax
:FUNCtion<m>:TRENd:NMEasurement?
The :FUNCtion<m>:TRENd:NMEasurement? query returns the selected
measurement.
If no measurements are installed, the :FUNCtion<m>:TRENd:NMEasurement?
query will return NONE.
Return Format
MEAS<n><NL>
<n> ::= # of installed measurement, from 1 to 8
See Also
•