Common (*) Commands
6
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
193
*OPT (Option Identification)
Query Syntax
*OPT?
The *OPT? query reports the options installed in the instrument. This query returns
a string that identifies the module and its software revision level.
Return Format
0,0,<license info>
<license info> ::= <All field>, <reserved>, <reserved>, <reserved>,
<reserved>, <Memory>, <Low Speed Serial>, <Automotive Serial>,
<reserved>, <Frequency Response Analysis>, <Power Measurements>,
<RS-232/UART Serial>, <Segmented Memory>, <Mask Test>, <reserved>,
<reserved>, <reserved>, <reserved>, <reserved>, <reserved>,
<reserved>, <reserved>, <reserved>, <reserved>, <Educator's Kit>,
<Waveform Generator>, <MIL-1553/ARINC 429 Serial>, <Extended Video>,
<Advanced Math>, <reserved>, <reserved>, <reserved>, <reserved>,
<Digital Voltmeter/Counter>, <reserved>, <reserved>, <reserved>,
<reserved>, <reserved>, <Remote Command Logging>, <reserved>,
<SENT Serial>, <CAN FD Serial>, <CXPI Serial>, <NFC Trigger>,
<reserved>, <reserved>, <reserved>, <Manchester/NRZ Serial>,
<USB PD Serial>, <reserved>,
<Automotive Software>, <General Purpose Software>,
<Aerospace Software>, <Power Supply Test Software>,
<reserved>, <Near Field Communications (NFC) Software>,
<Software Bundle>
<All field> ::= {0 | All}
<reserved> ::= 0
<Memory> ::= {0 | memMax}
<Low Speed Serial> ::= {0 | EMBD}
<Automotive Serial> ::= {0 | AUTO}
<Frequency Response Analysis> ::= {0 | FRA}
<Power Measurements> ::= {0 | PWR}
<RS-232/UART Serial> ::= {0 | COMP}
<Segmented Memory> ::= {0 | SGM}
<Mask Test> ::= {0 | MASK}
<Educator's Kit> ::= {0 | EDK}
<Waveform Generator> ::= {0 | WAVEGEN}
<MIL-1553/ARINC 429 Serial> ::= {0 | AERO}
<Extended Video> ::= {0 | VID}
<Advanced Math> ::= {0 | ADVMATH}
<Digital Voltmeter/Counter>> ::= {0 | DVMCTR}