208
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
7
Root (:) Commands
:DIGitize
[<source>[,..,<source
>]] (see
n/a
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> | FFT |
SBUS{1 | 2}}
<source> can be repeated up to 5
times
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
:HWEenable <n> (see
:HWEenable? (see
<n> ::= 16-bit integer in NR1
format
n/a
:HWERregister:CONDiti
<n> ::= 16-bit integer in NR1
format
n/a
:HWERegister[:EVENt]?
<n> ::= 16-bit integer in NR1
format
:MTEenable <n> (see
:MTEenable? (see
<n> ::= 16-bit integer in NR1
format
n/a
:MTERegister[:EVENt]?
<n> ::= 16-bit integer in NR1
format
:OPEE <n> (see
<n> ::= 15-bit integer in NR1
format
n/a
:OPERegister:CONDitio
<n> ::= 15-bit integer in NR1
format
n/a
:OPERegister[:EVENt]?
<n> ::= 15-bit integer in NR1
format
:OVLenable <mask>
(see
:OVLenable? (see
<mask> ::= 16-bit integer in NR1
format as shown:
Bit Weight Input
--- ------ ----------
10
1024
Ext Trigger Fault
9
512
Channel 4 Fault
8
256
Channel 3 Fault
7
128
Channel 2 Fault
6
64
Channel 1 Fault
4
16
Ext Trigger OVL
3
8
Channel 4 OVL
2
4
Channel 3 OVL
1
2
Channel 2 OVL
0
1
Channel 1 OVL
n/a
:OVLRegister? (see
<value> ::= integer in NR1
format. See OVLenable for <value>
Table 72
Root (:) Commands Summary (continued)
Command
Query
Options and Query Returns