:FUNCtion<m> Commands
17
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
379
:FUNCtion<m>:DISPlay
Command Syntax
:FUNCtion<m>:DISPlay <display>
<m> ::= 1 to (# math functions) in NR1 format
<display> ::= {{1 | ON} | {0 | OFF}}
The :FUNCtion<m>:DISPlay command turns the display of the function on or off.
When ON is selected, the function operates as specified by the other
:FUNCtion<m> commands.
One math function can be displayed at a time. If one math function is on and then
another math function is turned on, the first math function will be turned off.
When math functions are off, they are still calculated so that they can be
cascaded, that is, used as a source for a higher math function.
Query Syntax
:FUNCtion<m>:DISPlay?
The :FUNCtion<m>:DISPlay? query returns whether the function display is on or
off.
Return Format
<display><NL>
<display> ::= {1 | 0}
See Also
•
"Introduction to :FUNCtion<m> Commands"
•
•
•
•
•
NOTE
Automatic vertical scaling of the math function waveform occurs when the function display is
turned from off to on. If you wait for the operation to complete (with an *OPC? query for
example), then query the math functions's scale, you can see the vertical scaling that was
automatically determined.