770
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
27
:SBUS<n> Commands
:SBUS<n>:CAN:SIGNal:BAUDrate
Command Syntax
:SBUS<n>:CAN:SIGNal:BAUDrate <baudrate>
<baudrate> ::= integer from 10000 to 4000000 in 100 b/s increments,
or 5000000
The :SBUS<n>:CAN:SIGNal:BAUDrate command sets the standard baud rate of
the CAN signal from 10 kb/s to 4 Mb/s in 100 b/s increments. If you enter a baud
rate that is not divisible by 100 b/s, the baud rate is set to the nearest baud rate
divisible by 100 b/s.
You can also set the baud rate of the CAN signal to 5 Mb/s. Fractional baud rates
between 4 Mb/s and 5 Mb/s are not allowed.
If the baud rate you select does not match the system baud rate, false triggers
may occur.
Query Syntax
:SBUS<n>:CAN:SIGNal:BAUDrate?
The :SBUS<n>:CAN:SIGNal:BAUDrate? query returns the current CAN baud rate
setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 10000 to 4000000 in 100 b/s increments,
or 5000000
See Also
•
"Introduction to :TRIGger Commands"
•
•
•
":SBUS<n>:CAN:SIGNal:DEFinition"
•