1100
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
31
:TRIGger Commands
:TRIGger:GLITch:LEVel
Command Syntax
:TRIGger:GLITch:LEVel <level>[,<source>]
<level> ::= .75 x full-scale voltage from center screen in NR3 format
for internal triggers
<level> ::= ±(external range setting) in NR3 format
for external triggers
<source> ::= {CHANnel<n> | EXTernal}
<n> ::= 1 to (# analog channels) in NR1 format
The :TRIGger:GLITch:LEVel command sets the trigger level voltage for the active
pulse width trigger.
Query Syntax
:TRIGger:GLITch:LEVel?
The :TRIGger:GLITch:LEVel? query returns the trigger level of the current pulse
width trigger mode. If all channels are off, the query returns "NONE."
Return Format
<level><NL>
See Also
•
"Introduction to :TRIGger Commands"
•
•
•