1336
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
37
Status Reporting
Mask Test Event Event Register (:MTERegister[:EVENt])
The Mask Test Event Event Register register hosts these bits:
The :MTERegister[:EVENt]? query returns the value of, and clears, the Mask Test
Event Event Register.
See Also
•
":MTERegister[:EVENt] (Mask Test Event Event Register)"
Name
Location
Description
Complete
bit 0
Is set when the mask test is complete.
Fail
bit 1
Is set when there is a mask test failure.
Started
bit 8
Is set when mask testing is started.
Pass
bit 9
Is set when the mask test passes.
Auto Mask
bit 10
Is set when auto mask creation is completed.