:POWer Commands
24
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
683
:POWer:SWITch:IREFerence
Command Syntax
:POWer:SWITch:IREFerence <percent>
<percent> ::= percent in NR1 format
The :POWer:SWITch:IREFerence command to specify the current switching level
for the start of switching edges. The value is in percentage of the maximum switch
current.
You can adjust this value to ignore noise floors or null offset that is difficult to
eliminate in current probes.
This value specifies the threshold that is used to determine the switching edges.
Query Syntax
:POWer:SWITch:IREFerence?
The :POWer:SWITch:IREFerence query returns the current switching level percent
value.
Return Format
<percent><NL>
<percent> ::= percent in NR1 format
See Also
•
•
•
•
•