Obsolete and Discontinued Commands
35
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1289
:MEASure:VDELta
Query Syntax
:MEASure:VDELta?
The :MEASure:VDELta? query returns the voltage difference between vertical
marker 1 (Y1 cursor) and vertical marker 2 (Y2 cursor). No measurement is made
when the :MEASure:VDELta? query is received by the oscilloscope. The delta value
that is returned is the current value. This is the same value as the front-panel
cursors delta Y value.
VDELta = value at marker 2 - value at marker 1
Return Format
<value><NL>
<value> ::= delta V value in NR1 format
See Also
•
"Introduction to :MARKer Commands"
•
"Introduction to :MEASure Commands"
•
•
•
•
•
NOTE
The :MEASure:VDELta command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:YDELta command (see
) instead.