:TRIGger Commands
31
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1121
:TRIGger:PATTern:FORMat
Command Syntax
:TRIGger:PATTern:FORMat <base>
<base> ::= {ASCii | HEX}
The :TRIGger:PATTern:FORMat command sets the entry (and query) number base
used by the :TRIGger:PATTern command. The default <base> is ASCii.
Query Syntax
:TRIGger:PATTern:FORMat?
The :TRIGger:PATTern:FORMat? query returns the currently set number base for
pattern trigger patterns.
Return Format
<base><NL>
<base> ::= {ASC | HEX}
See Also
•
"Introduction to :TRIGger Commands"
•