1246
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
33
:WGEN<w> Commands
:WGEN<w>:VOLTage
Command Syntax
:WGEN<w>:VOLTage <amplitude>
<w> ::= 1 to (# WaveGen outputs) in NR1 format
<amplitude> ::= amplitude in volts in NR3 format
For all waveforms except DC, the :WGEN<w>:VOLTage command specifies the
waveform's amplitude. Use the :WGEN<w>:VOLTage:OFFSet command to specify
the offset voltage or DC level.
You can also specify the amplitude and offset indirectly using the
:WGEN<w>:VOLTage:HIGH and :WGEN<w>:VOLTage:LOW commands. For
example, an amplitude of 5 V and an offset of 1 V is the same as a high-level
voltage of 4 V and a low-level voltage of -1 V.
Query Syntax
:WGEN<w>:VOLTage?
The :WGEN<w>:VOLTage? query returns the currently specified waveform
amplitude.
Return Format
<amplitude><NL>
<amplitude> ::= amplitude in volts in NR3 format
See Also
•
"Introduction to :WGEN<w> Commands"
•
•
•
•
NOTE
When the amplitude is below 40 mV, the offset is limited to ±500 mV.