:CHANnel<n> Commands
10
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
291
:CHANnel<n>:OFFSet
Command Syntax
:CHANnel<n>:OFFSet <offset> [<suffix>]
<offset> ::= Vertical offset value in NR3 format
<suffix> ::= {V | mV}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:OFFSet command sets the value that is represented at center
screen for the selected channel. The range of legal values varies with the value set
by the :CHANnel<n>:RANGe and :CHANnel<n>:SCALe commands. If you set the
offset to a value outside of the legal range, the offset value is automatically set to
the nearest legal value. Legal values are affected by the probe attenuation setting.
Query Syntax
:CHANnel<n>:OFFSet?
The :CHANnel<n>:OFFSet? query returns the current offset value for the selected
channel.
Return Format
<offset><NL>
<offset> ::= Vertical offset value in NR3 format
See Also
•
"Introduction to :CHANnel<n> Commands"
•
•
•