Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
24
:POWer Commands
:POWer:CLResponse:WGEN:LOAD
Command Syntax
:POWer:CLResponse:WGEN:LOAD <impedance>
<impedance> ::= {ONEMeg | FIFTy}
The :POWer:CLResponse:WGEN:LOAD command sets the waveform generator
expected output load impedance.
The output impedance of the Gen Out signal is fixed at 50 ohms. However, the
output load selection lets the waveform generator display the correct amplitude
and offset levels for the expected output load. If the actual load impedance is
different than the selected value, the displayed amplitude and offset levels will be
incorrect.
Query Syntax
:POWer:CLResponse:WGEN:LOAD?
The :POWer:CLResponse:WGEN:LOAD? query returns the waveform generator
expected output load impedance setting.
Return Format
<impedance><NL>
<impedance> ::= {ONEM | FIFT}
See Also
•
•
•
•
":POWer:CLResponse:FREQuency:MODE"
•
":POWer:CLResponse:FREQuency:STARt"
•
":POWer:CLResponse:FREQuency:STOP"
•
•
":POWer:CLResponse:SOURce:INPut"
•
":POWer:CLResponse:SOURce:OUTPut"
•
":POWer:CLResponse:WGEN:VOLTage"
•