1152
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
31
:TRIGger Commands
:TRIGger:TV:MODE
Command Syntax
:TRIGger:TV:MODE <mode>
<mode> ::= {FIEld1 | FIEld2 | AFIelds | ALINes | LINE | LFIeld1
| LFIeld2 | LALTernate}
The :TRIGger:TV:MODE command selects the TV trigger mode and field. The
LALTernate parameter is not available when :TRIGger:TV:STANdard is GENeric.
Old forms for <mode> are accepted:
Query Syntax
:TRIGger:TV:MODE?
The :TRIGger:TV:MODE? query returns the TV trigger mode.
Return Format
<value><NL>
<value> ::= {FIE1 | FIE2 | AFI | ALIN | LINE | LFI1 | LFI2 | LALT}
See Also
•
"Introduction to :TRIGger Commands"
•
•
<mode>
Old Forms Accepted
FIEld1
F1
FIEld2
F2
AFIelds
ALLFields, ALLFLDS
ALINes
ALLLines
LFIeld1
LINEF1, LINEFIELD1
LFIeld2
LINEF2, LINEFIELD2
LALTernate
LINEAlt