Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
10
:CHANnel<n> Commands
:CHANnel<n>:SCALe
Command Syntax
:CHANnel<n>:SCALe <scale>[<suffix>]
<scale> ::= vertical units per division in NR3 format
<suffix> ::= {V | mV}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:SCALe command sets the vertical scale, or units per division, of
the selected channel.
If the probe attenuation is changed, the scale value is multiplied by the probe's
attenuation factor.
Query Syntax
:CHANnel<n>:SCALe?
The :CHANnel<n>:SCALe? query returns the current scale setting for the specified
channel.
Return Format
<scale value><NL>
<scale value> ::= vertical units per division in NR3 format
See Also
•
"Introduction to :CHANnel<n> Commands"
•
•