:TRIGger Commands
31
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1155
:TRIGger:TV:STANdard
Command Syntax
:TRIGger:TV:STANdard <standard>
<standard> ::= {GENeric | NTSC | PALM | PAL | SECam
| {P480L60HZ | P480} | {P720L60HZ | P720}
| {P1080L24HZ | P1080} | P1080L25HZ
| P1080L50HZ | P1080L60HZ
| {I1080L50HZ | I1080} | I1080L60HZ}
The :TRIGger:TV:STANdard command selects the video standard:
• NTSC
• PAL
• PAL-M
• SECAM
With an extended Video triggering license, the oscilloscope additionally supports
these standards:
• Generic — GENeric mode is non-interlaced.
• EDTV 480p/60
• HDTV 720p/60
• HDTV 1080p/24
• HDTV 1080p/25
• HDTV 1080i/50
• HDTV 1080i/60
Query Syntax
:TRIGger:TV:STANdard?
The :TRIGger:TV:STANdard? query returns the current TV trigger standard setting.
Return Format
<standard><NL>
<standard> ::= {GEN | NTSC | PALM | PAL | SEC | P480L60HZ | P760L60HZ
| P1080L24HZ | P1080L25HZ | P1080L50HZ | P1080L60HZ
| I1080L50HZ | I1080L60HZ}