:TRIGger Commands
31
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1091
:TRIGger[:EDGE] Commands
Table 132
:TRIGger[:EDGE] Commands Summary
Command
Query
Options and Query Returns
:TRIGger[:EDGE]:COUPl
ing {AC | DC |
LFReject} (see
)
:TRIGger[:EDGE]:COUPl
{AC | DC | LFReject}
:TRIGger[:EDGE]:LEVel
<level> [,<source>]
(see
)
:TRIGger[:EDGE]:LEVel
? [<source>] (see
For internal triggers, <level>
::= .75 x full-scale voltage from
center screen in NR3 format.
For external triggers, <level>
::= ±(external range setting) in
NR3 format.
<source> ::= {CHANnel<n> |
EXTernal}
<n> ::= 1 to (# analog channels)
in NR1 format
:TRIGger[:EDGE]:REJec
t {OFF | LFReject |
HFReject} (see
)
:TRIGger[:EDGE]:REJec
{OFF | LFReject | HFReject}
:TRIGger[:EDGE]:SLOPe
<polarity> (see
)
:TRIGger[:EDGE]:SLOPe
<polarity> ::= {POSitive |
NEGative | EITHer | ALTernate}
:TRIGger[:EDGE]:SOURc
e <source> (see
)
:TRIGger[:EDGE]:SOURc
<source> ::= {CHANnel<n> |
EXTernal | WGEN | WGEN1 | WMOD}
<n> ::= 1 to (# analog channels)
in NR1 format
Note: WGEN and WGEN1 are
equivalent.