1302
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
35
Obsolete and Discontinued Commands
:SBUS<n>:LIN:SIGNal:DEFinition
Command Syntax
:SBUS<n>:LIN:SIGNal:DEFinition <value>
<value> ::= {LIN | RX | TX}
The :SBUS<n>:LIN:SIGNal:DEFinition command sets the LIN signal type. These
signals can be set to:
Dominant low signals:
• LIN — the actual LIN single-end bus signal line.
• RX — the Receive signal from the LIN bus transceiver.
• TX — the Transmit signal to the LIN bus transceiver.
Query Syntax
:SBUS<n>:LIN:SIGNal:DEFinition?
The :SBUS<n>:LIN:SIGNal:DEFinition? query returns the current LIN signal type.
Return Format
<value><NL>
<value> ::= LIN
See Also
•
"Introduction to :TRIGger Commands"
•
•
":SBUS<n>:LIN:SIGNal:BAUDrate"
•
NOTE
This command is available, but the only legal value is LIN.