:TRIGger Commands
31
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1099
:TRIGger:GLITch:LESSthan
Command Syntax
:TRIGger:GLITch:LESSthan <less_than_time>[<suffix>]
<less_than_time> ::= floating-point number in NR3 format
<suffix> ::= {s | ms | us | ns | ps}
The :TRIGger:GLITch:LESSthan command sets the maximum pulse width duration
for the selected :TRIGger:GLITch:SOURce.
Query Syntax
:TRIGger:GLITch:LESSthan?
The :TRIGger:GLITch:LESSthan? query returns the pulse width duration time for
:TRIGger:GLITch:SOURce.
Return Format
<less_than_time><NL>
<less_than_time> ::= floating-point number in NR3 format.
See Also
•
"Introduction to :TRIGger Commands"
•
•
•