1106
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
31
:TRIGger Commands
:TRIGger:NFC:AEVent
Command Syntax
:TRIGger:NFC:AEVent <arm_event>
<arm_event> ::= {NONE | ASReq | AALLreq | AEITher | BSReq | BALLreq
| BEITher | FSReq}
When the ATRigger (Arm & Trigger) trigger event is selected (by
:TRIGger:NFC:TEVent), the :TRIGger:NFC:AEVent command specifies the arm
event. Valid arm event settings depend on the signaling technology selected (by
:TRIGger:NFC:STANdard):
When a trigger event other than ATRigger (Arm & Trigger) is selected, the arm
event is NONE.
Query Syntax
:TRIGger:NFC:AEVent?
The :TRIGger:NFC:AEVent? query returns the specified arm event.
Return Format
<arm_event><NL>
<arm_event> ::= {NONE | ASR | AALL | AEIT | BSR | BALL | BEIT | FSR}
See Also
•
•
•
•
•
•
•
•
Signaling
Technology
Arm Event
Description
NFC-A
ASReq
SENS_REQ
AALLreq
ALL_REQ
AEITher
Either the SENS_REQ or ALL_REQ events will arm.
NFC-B
BSReq
SENSB_REQ
BALLreq
ALLB_REQ
BEITher
Either the SENSB_REQ or ALLB_REQ events will arm.
NFC-F
FSReq
SENSF_REQ