600
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
24
:POWer Commands
:POWer:CLResponse:DATA:GMARgin
Query Syntax
:POWer:CLResponse:DATA:GMARgin?
After the Control Loop Response (Bode) power analysis has been performed (see
:POWer:CLResponse:APPLy), the :POWer:CLResponse:DATA:GMARgin? query
returns the gain margin in dB.
Return Format
<gain_margin><NL>
<gain_margin> ::= gain margin in dB in NR3 format.
The query r9.9E+37 if the value cannot be calculated from the last sweep
(that is, if there are no crossover points).
See Also
•
•
•
":POWer:CLResponse:DATA:GMARgin:FREQuency"
•
":POWer:CLResponse:DATA:PMARgin"
•