background image

Made in the USA 

 

 

 

 
 

MDS

 

 

Mobile Display System

 

 

 

 

Installation & Operator 

Manual 

Summary of Contents for MDS

Page 1: ...Made in the USA MDS Mobile Display System Installation Operator Manual...

Page 2: ...nd or international agencies regarding specific requirements and regulations applicable to the use of this equipment Image Diagnostics Inc certifies only the equipment After sale operating practices a...

Page 3: ...ALLATION INSTRUCTIONS 6 6 1 Mount Two Main Monitors 7 6 2 Release Monitor Counterbalance 9 6 3 Shelf Adjustment 10 6 4 Clean Unit 11 6 5 Electronics Installation 11 7 OPERATOR INSTRUCTIONS 12 7 1 Movi...

Page 4: ...eath personal injury or damage to the equipment Recyclable material Protective Ground This is the common tie point between the AC Electrical Power Cord Ground Frame Ground and Controller Ground SGS No...

Page 5: ...tand its operation instructions emergency procedures capabilities including total range of motion and be aware of all potential safety hazards This manual should be accessible to all personnel install...

Page 6: ...utlet not to exceed the rated amp values as stated on the MDS serial number label Safety Hazard Level Potential Consequences with Use DANGER Indicates an imminently hazardous situation which if not av...

Page 7: ...operator inadvertently runs into the extended monitor arms Be aware of the arm s location at all times during the installation and while in use CAUTION Prior to each clinical placement confirm that t...

Page 8: ...nternal components may result in increased emissions or decreased immunity of the equipment or system WARNING If MDS is used adjacent to or stacked with other medical electrical equipment the equipmen...

Page 9: ...oks should ONLY be utilized during transportation of the unit Note The original factory balance weight is noted on the face of the VESA Monitor Support Bracket and the Face Balance Plates If Monitors...

Page 10: ...s are secure and in place against the top and bottom Counterweight Bracket Flanges as shown in the photos below NOTE The upper and lower Locking Collars are used to secure the Arms for shipping only a...

Page 11: ...6 With assistance from another individual lift the Monitor with the partially installed upper screws and slide the screws into the upper slots of the VESA Mounting Plates on the MDS arm 6 1 7 Apply 22...

Page 12: ...d secure the counterweight Locking Collars are ready to be relocated from their storage positions 6 2 2 With a 3 16 Hex Key loosen the lower Locking Collars and lower them to the bottom of the slide r...

Page 13: ...rvice for procedure steps NOTE There are inner and outer locking collars for each balance plate 6 3 Shelf Adjustment Tools 3 16 Hex Key a Level and an assistant Procedure 6 3 1 Remove all devices off...

Page 14: ...ite generic household bleach in a solution of 5 25 sodium hypochlorite diluted between 1 10 and 1 100 with water o Alcohol generic o Envirocide Procedure 6 4 1 Unplug the MDS system prior to cleaning...

Page 15: ...re pushed in and Magnets are engaged 7 1 4 Lift the arms so that the base of the monitor clears the upper most Device placed on the Top Shelf and rotate the face of one of the Monitors inward and gent...

Page 16: ...r 7 2 2 Push and pull monitor to desired location within the limits of the articulating arms they are mounted to 7 2 3 To tilt monitor position downward hold the top of the monitor mount and push the...

Page 17: ...manual and the proper tools Lubrication of this device is not required 8 1 Recommended Periodic Performance Checks Daily Inspect all external cables and electrical components for wear and damage Damag...

Page 18: ...e appropriate lockout tag out procedure while performing any diagnostics and service on the unit 9 TROUBLESHOOTING Problem Symptom Possible Cause Remedy 1 MDS generates or receives electrical interfer...

Page 19: ...t to 50 A Isolation transformer is certified to meet Medical Safety Standards ANSI AAMI ES60601 1 Environmental Operating Temperature Range 41 F to 95 F 5 C to 35 C Operating Humidity Range 10 to 85 r...

Page 20: ...stics com 17 12 WARRANTY Warranty details for IDI Products can be obtained directly from Image Diagnostics Inc Image Diagnostics Inc 310 Authority Drive Fitchburg MA 01420 USA Or call IDI at 978 829 0...

Reviews: