1234
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
33
:WGEN<w> Commands
:WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry
Command Syntax
:WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry <percent>
<w> ::= 1 in NR1 format
<percent> ::= symmetry percentage from 0% to 100% in NR1 format
The :WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry command specifies the
amount of time per cycle that the ramp waveform is rising. The ramp modulating
waveform shape is specified with the :WGEN<w>:MODulation:FUNCtion
command.
Query Syntax
:WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry?
The :WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry? query returns ramp
symmetry percentage setting.
Return Format
<percent><NL>
<percent> ::= symmetry percentage from 0% to 100% in NR1 format
See Also
•
":WGEN<w>:MODulation:AM:DEPTh"
•
":WGEN<w>:MODulation:AM:FREQuency"
•
":WGEN<w>:MODulation:FM:DEViation"
•
":WGEN<w>:MODulation:FM:FREQuency"
•
":WGEN<w>:MODulation:FSKey:FREQuency"
•
":WGEN<w>:MODulation:FSKey:RATE"
•
":WGEN<w>:MODulation:FUNCtion"
•
•