:WGEN<w> Commands
33
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1233
:WGEN<w>:MODulation:FUNCtion
Command Syntax
:WGEN<w>:MODulation:FUNCtion <shape>
<w> ::= 1 in NR1 format
<shape> ::= {SINusoid | SQUare| RAMP}
The :WGEN<w>:MODulation:FUNCtion command specifies the shape of the
modulating signal.
When the RAMP shape is selected, you can specify the amount of time per cycle
that the ramp waveform is rising with the
:WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry command.
This command applies to AM and FM modulation. (The FSK modulation signal is a
square wave shape.)
Query Syntax
:WGEN<w>:MODulation:FUNCtion?
The :WGEN<w>:MODulation:FUNCtion? query returns the specified modulating
signal shape.
Return Format
<shape><NL>
<shape> ::= {SIN | SQU| RAMP}
See Also
•
":WGEN<w>:MODulation:AM:DEPTh"
•
":WGEN<w>:MODulation:AM:FREQuency"
•
":WGEN<w>:MODulation:FM:DEViation"
•
":WGEN<w>:MODulation:FM:FREQuency"
•
":WGEN<w>:MODulation:FSKey:FREQuency"
•
":WGEN<w>:MODulation:FSKey:RATE"
•
":WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry"
•
•