:POWer Commands
:POWer:HARMonics:STANdard
Command Syntax
:POWer:HARMonics:STANdard <class>
<class> ::= {A | B | C | D}
The :POWer:HARMonics:STANdard command selects the standard to perform
current harmonics compliance testing on.
• A — IEC 61000-3-2 Class A — for balanced three-phase equipment, household
appliances (except equipment identified as Class D), tools excluding portable
tools, dimmers for incandescent lamps, and audio equipment.
• B — IEC 61000-3-2 Class B — for portable tools.
• C — IEC 61000-3-2 Class C — for lighting equipment.
• D — IEC 61000-3-2 Class D — for equipment having a specified power
according less than or equal to 600 W, of the following types: personal
computers and personal computer monitors, television receivers.
Query Syntax
:POWer:HARMonics:STANdard?
The :POWer:HARMonics:STANdard query returns the currently set IEC 61000-3-2
standard.
Return Format
<class><NL>
<class> ::= {A | B | C | D}
See Also
•
•
•
•
•
•
":POWer:HARMonics:POWerfactor"
•
•
•