:POWer Commands
24
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
649
:POWer:PSRR:FREQuency:MODE
Command Syntax
:POWer:PSRR:FREQuency:MODE <mode>
<mode> ::= {SWEep | SINGle}
The :POWer:PSRR:FREQuency:MODE command specifies whether the analysis
should be performed by sweeping through a range of frequencies (SWEep) or at a
single frequency (SINGle).
The SINGle mode is useful for evaluating amplitudes at a single frequency. After
running the test at a single frequency, you can manually adjust (increase) the
waveform generator's amplitude until you begin to observe distortion in the
waveforms on the oscilloscope's display. You can then use that amplitude at all
frequencies in SWEep mode, or you can evaluate amplitudes at other frequencies
in order to determine an optimized amplitude profile (see
:POWer:PSRR:WGEN:VOLTage:PROFile).
Query Syntax
:POWer:CLResponse:FREQuency:MODE?
The :POWer:PSRR:FREQuency:MODE? query returns the frequency mode setting.
Return Format
<mode><NL>
<mode> ::= {SWE | SING}
See Also
•
•
•
•
":POWer:PSRR:FREQuency:MAXimum"
•
":POWer:PSRR:FREQuency:MINimum"
•
":POWer:PSRR:FREQuency:SINGle"
•
•
•
•
•
•