Obsolete and Discontinued Commands
35
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1291
:MEASure:VSTOp
Command Syntax
:MEASure:VSTOp <vstop_argument>
<vstop_argument> ::= value for Y2 cursor
The :MEASure:VSTOp command moves the vertical marker 2 (Y2 cursor) to the
specified value corresponding to the selected source. The source can be selected
by the MARKer:X2Y2source command.
Query Syntax
:MEASure:VSTOp?
The :MEASure:VSTOp? query returns the current value of the Y2 cursor.
Return Format
<value><NL>
<value> ::= value of the Y2 cursor in NR3 format
See Also
•
"Introduction to :MARKer Commands"
•
"Introduction to :MEASure Commands"
•
•
•
•
•
•
•
NOTE
The short form of this command, VSTO, does not follow the defined Long Form to Short Form
Truncation Rules (see
). The normal short form, VST, would be the same for both
VSTArt and VSTOp, so sending VST for the VSTOp command produces an error.
NOTE
The :MEASure:VSTOp command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:Y2Position command (see
) instead.