background image

 

 

 

 

 

 
 
 
 
 
 

 

FreeStyle

®   

Comfort

 

 

 

 

 
 

Technical Manual 

(English) MN233 B 

 
 
 
 

Summary of Contents for FreeStyle Comfort

Page 1: ...FreeStyle Comfort Technical Manual English MN233 B...

Page 2: ...from open flame fire sparks Open ignition source and smoking prohibited Reg P003 Type BF applied part degree of protection against electric shock Reg 5333 Warning Reg W001 Council Directive 93 42 EEC...

Page 3: ...vice meets the requirements for modular transmitter approval as detailed in FCC public Notice DA00 1407 Transmitter operation is subject to the following two conditions 1 This device may not cause har...

Page 4: ...iew 5 Description of Operation 7 Safety Instructions 8 Specifications 13 Test Procedures 14 Patient Cleaning Care Routine Maintenance 19 Equipment Provider Maintenance 21 Service Procedures 24 Alarm C...

Page 5: ...omfort Overview Front View Strap Attachment User Interface Air Outlet Oxygen Outlet Battery Pack Air Intake Filter Battery Pack Release Button Power Port Receptacle Tab Decrease Setting Power Button I...

Page 6: ...6 FreeStyle Comfort Display Screen on Battery Display Screen On AC DC Power...

Page 7: ...ivered to the user The FreeStyle Comfort combines advanced oxygen concentrator and oxygen conserving technologies to create a portable device that allows for maximum portability and ambulation The adv...

Page 8: ...INTENDED FOR USE WITH A TRACHEOTOMISED PATIENT WARNING PREGNANT OR NURSING WOMEN SHOULD NOT USE ACCESSORIES RECOMMENDED IN THIS MANUAL THEY MAY CONTAIN PHTHALATES WARNING IN CERTAIN CIRCUM STANCES THE...

Page 9: ...TEMPERATURE LIMITS WARNING USE NO OIL GREASE OR PETRO LEUM BASED OR OTHER FLAMMABLE PROD UCTS WITH THE OXYGEN CARRYING ACCESSO RIES OR THE OXYGEN CONCENTRATOR ONLY WATER BASED OXYGEN COMPATIBLE LOTION...

Page 10: ...G DO NOT ALLOW SMOKING CAN DLES OR OPEN FLAMES IN THE SAME ROOM WITH THE DEVICE OR THE OXYGEN CARRYING ACCESSORIES WARNING SMOKING WHILE WEARING AN OXYGEN CANNULA MAY CAUSE FACIAL BURNS AND POSSIBLY D...

Page 11: ...m children CAUTION ONLY USE the Manufacturer s provided batteries For proper battery disposal contact your Equipment Provider or your local government agency for disposal requirements CAUTION Operatin...

Page 12: ...REMOVE THE BATTERY PACK WHILE THE UNIT IS STILL ON OR PLUGGED INTO THE POWER SUPPLY Note If the oxygen concentrator has been stored for an extended period of time outside of its normal environmental...

Page 13: ...uble Battery Duration Setting 2 up to 8 hours Operating Temperature 41 F to 104 F 5 C to 40 C Transportation and Storage Temp Unit 13 F to 158 F 25 C to 70 C up to 10 000 ft 3048m Transportation and S...

Page 14: ...ttery with the bottom housing from the center of the FreeStyle Comfort b Slide the battery inward until you hear a click to confirm that it has locked in place 3 Insert the AC or DC power supply into...

Page 15: ...e When the Freestyle Comfort is initially turned on the default flow setting is 2 If the setting is changed the unit will power on to the previous flow setting Note The display screen will enter stand...

Page 16: ...testing such as oxygen level 1 To check the Alarm Code turn the unit on and enter the Test Mode per instruct from the previous section 2 On the top of the screen the four digit number indicates the o...

Page 17: ...ector inlet on the front right bottom of the FreeStyle Comfort Be sure to push inward so that the connection is secure 5 Turn on the FreeStyle Comfort by pressing the power button and increase or decr...

Page 18: ...er source is connected If an electrical outlet is available operable and its location is appropriate for use as a power source it may serve as a back up to the FreeStyle Comfort s Batteries In this ca...

Page 19: ...d water 3 Rinse the filters thoroughly and remove excess water with a soft lint free cloth Ensure thefilters are completely dry before reattaching 4 Re install the clean dry filters BATTERY PACK CARE...

Page 20: ...cabinet and power supplies Proceed as directed by the cleaner manufacturer WARNING TURN OFF THE UNIT AND DISCONNECT THE POWER CORD FROM THE ELECTRICAL OUTLET BEFORE YOU CLEAN THE UNIT TO PREVENT ACCI...

Page 21: ...ructions below 1 Turn the unit off disconnect the power cord and remove the power supply 2 While using the oxygen outlet Spanner and a nut driver unscrew oxygen outlet on the front top right of the un...

Page 22: ...replace the air intake filter with warm soapy water between each patient use The patients should be instructed to clean this filter at least once per week depending on the environment during normal op...

Page 23: ...is performed on a unit an entry should be made in the service log for that concentrator or recorded in accordance with your company s standard procedure FreeStyle Comfort Serial Number DATE HOUR METE...

Page 24: ...of the sieve bed 5 Flip the unit in the upright position and pull down on the sieve beds to remove from the manifold 6 Perform the steps in reverse to replace the old sieve beds with new sieve beds WA...

Page 25: ...en internal temperature is approaching shutdown threshold If shutdown threshold is reached unit will shut down and alarm screen will display and audible alarm will continue for 2 min 3 System Malfunct...

Page 26: ...sing until a breath is detected or Warm up ends whichever is sooner 7 Low concentration Alarm 1 beep every 30 sec for up to 5 min 8 System Malfunction Flow rate OCD Valve Failure Obstruction of Gas Pa...

Page 27: ...ressure Alarm None 13 System Malfunction Low Temp Alarm None NA Airplane Mode Info None Airplane Mode on battery power NA Airplane Mode Info None Airplane Mode on external power All visual alarms are...

Page 28: ...the compressor following the air flow of the unit to the oxygen outlet Repair all leaks by tightening connections and fittings 5 To exit test mode turn off the unit Be sure to exit test mode before r...

Page 29: ...ery 0x0020 O2 Warmup The unit is Warming up Allow the unit to warm up for 15 minutes 0x0040 Low O2 Intake Filter Clogged Sieve Bed Failure Low pressure High pressure Other Clean or replace intake filt...

Page 30: ...Clogged Pinched or Clogged Inlet Tubing Feed Waste Valve Failure Sieve Bed Failure Other Clean or replace intake filters Verify that compressor inlet tubing is not pinched or clogged Replace feed wast...

Page 31: ...ery 8 Cell BT034 1S Battery 16 Cell BT035 1S Soft Bag MI396 1 Shoulder Strap MI397 1 Power Supply 120W 15V PW036 1S Oxygen Outlet Spanner 21168706 DC Cord CD041 1 Optional Accessories Available Deskto...

Page 32: ...32 FreeStyle Comfort Notes...

Page 33: ...o mps gmbh eu Worldwide www CAIREmedical com Copyright 2018 Chart Industries CAIRE Inc reserves the right to discontinue its products or change the prices materials equipment quality descriptions spec...

Reviews: