:SBUS<n> Commands
27
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
789
:SBUS<n>:CXPI:BAUDrate
Command Syntax
:SBUS<n>:CXPI:BAUDrate <baudrate>
<baudrate> ::= integer from 9600 to 40000 in 100 b/s increments.
The :SBUS<n>:CXPI:BAUDrate command specifies the baud rate of the CXPI
signal from your device under test.
The CXPI baud rate can be set from 9600 b/s to 40000 b/s in 100 b/s increments.
You must set the baud rate to match your device under test.
The default baud rate is 20 kb/s.
Query Syntax
:SBUS<n>:CXPI:BAUDrate?
The :SBUS<n>:CXPI:BAUDrate? query returns the baud rate setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 9600 to 40000 in 100 b/s increments.