background image

PL-0370 Rev 00 / December 2019 / Page 1

ID

Alarm/

Alert

Text Mes-

sage

System Status & OperatorResponse

S1

Alarm

POWER ON 

TEST FAIL

The Pump will not start. An audible alarm will sound, which 

cannot be muted. Switch the Console OFF and ON again. If 

the alarm reappears, switch to the backup Console, Motor 

and Flow Probe, record the alarm message and contact 

your local Abbott Medical representative.

S2

Alarm

SYSTEM 

FAULT

(Run-Time 

System 

Failure)

The Pump will stop. An audible alarm will sound, which 

cannot be muted. Clamp the return tubing and switch to 

the backup Console, Motor and Flow Probe according the 

procedure described in Section 2. Resume support. Record 

the alarm message and contact your local Abbott Medical 

representative.

M1

Alarm

MOTOR 

STOPPED

The Pump will stop. An audible alarm will sound, which 

can be muted for 60 seconds.Clamp the return tubing 

and switch to the backup Console, Motor and Flow Probe 

according the procedure described in Section 2. Resume 

support. Record the alarm message and contact your local 

Abbott Medical representative.

M2

Alarm

MOTOR 

DISCON-

NECTED

The Pump will stop. An audible alarm will sound, which can 

be muted for 60 seconds.

During setup of the System:

Press the alarm acknowledge button and check that the Mo-

tor connector is fully inserted into the back of the Console. 

During support:

Press the alarm acknowledge button and check that the 

connector of the Motor is fully inserted into the back of the 

Console. Resume support. If the visual alarm message 

does not disappear, clamp the return tubing and switch to 

the backup Console, Motor and Flow Probe according the 

procedure described in Section 2. Resume support. Record 

the alarm message and contact your local Abbott Medical 

representative.

M3

Alarm

PUMP NOT 

INSERTED

System will not start. An audible alarm will sound, which can 

be muted for 60 seconds. Press the alarm acknowledge 

button. Insert or reinsert the pump and secure it with 

the locking screw. If the alarm repeats, switch to backup 

Console, Motor and Flow Probe according the procedure 

described in Section 2.

M4

Alarm

MOTOR 

ALARM

An audible alarm will sound and the System will continue 

to operate. Press the alarm acknowledge button, if the 

visual alarm message does not disappear, clamp the return 

tubing, stop the pump and switch to the backup Console, 

Motor and Flow Probe according the procedure described 

in Section 2. Resume support. Record the alarm message 

and contact your local Abbott Medical representative.

M5

Alert

SET PUMP 

SPEED NOT 

REACHED

Check the Pump flow.

 If pump flow is satisfactory,

 reduce 

set  speed  while  insuring  that  flow  is  maintained.  Press 

the alarm acknowledge button. If the alert repeats, clamp 

the return tubing, stop the pump and switch to the backup 

Console, Motor and Flow Probe according the procedure 

described in Section 2. Resume support. Record the alert 

message and contact your Abbott Medical representative.

If the pump flow is not satisfactory

, clamp the return tubing 

and switch to the backup Console, Motor and Flow Probe 

according to the procedure described in Section 2. Resume 

support. Record the alert message and contact Abbott Med-

ical representative.

ID

Alarm/

Alert

Text Mes-

sage

System Status & OperatorResponse

B1

Alert

BATTERY 

MODULE 

FAIL

The Console battery will not function. An audible alarm 

will sound. Switch to the backup Console, Motor and Flow 

Probe according to the procedure described in Section 2. 

Resume support. Record the alert message and contact 

your Abbott Medical representative.

B2

Alert

BATTERY 

BELOW 

MINIMUM

The Pump will stop after a very short time. Plug the Console 

into a AC power outlet to charge the battery. If no AC outlet 

is available, switch to the backup Console, Motor and Flow 

Probe according to the procedure described in Section 2.

Resume support.

F1

Alert

FLOW 

PROBE 

DISCON-

NECTED

Check the Flow Probe connection on back of the Console. 

If necessary, reconnect the Flow Probe connector to the 

back of the Console. Press the alarm acknowledge button. 

Switch to the backup Flow Probe, if the alert message 

repeats.

S3

Alert

SYSTEM 

ALERT

Press the alarm acknowledge button, if the message does 

not disappear, clamp the return tubing, stop the pump 

and switch to the backup Console, Motor and Flow Probe 

according the procedure described in Section 2. Resume 

support. Record the alert message and contact your local 

Abbott Medical representative.

F2

Alert

FLOW

SIGNAL

INTER-

RUPTED

(Flow rate 

sensor error)

Manually disconnect, reposition and reconnect the 

Flow Probe transducer to the tubing. Press the alarm 

acknowledge button. Switch to the backup Flow Probe, 

if the alert message repeats. If problem still persists after 

switching to the backup Flow Probe, stop the pump and 

switch to a backup Console. Follow the instructions 

described in Section 2. Resume support. Record the alert 

message and contact your Abbott Medical representative.

F3

Alert

FLOW

BELOW

MINIMUM

(Low Flow)

Check for physiologic cause or circuit obstruction.

Check  minimum  flow  set  point.  Do  not  increase  RPM 

without confirming that adequate blood volume is available. 

Common cause of this alert is inadequate blood volume at 

the drainage cannula site for the desired pump flow.

F4

Alert

FLOW 

ABOVE 

MAXIMUM

Reduce Pump speed and check for cause.

P1

Alert

PRESSURE 

1 DISCON-

NECTED

Check the electrical connections on the pressure 1 trans-

ducer and recalibrate. If the problem persists disconnect, 

reconnect, and recalibrate the transducer.  Consider chang-

ing the transducer and cable if the problem persists.

P2

Alert

PRESSURE 

2 DISCON-

NECTED

Check the electrical connections on the pressure 2 trans-

ducer and recalibrate. If the problem persists disconnect, 

reconnect, and recalibrate the transducer.  Consider chang-

ing the transducer and cable if the problem persists.

ID

Alarm/

Alert

Text Mes-

sage

System Status & OperatorResponse

P3

Alert

PRESSURE 

SYSTEM 

FAIL

The pressure monitoring system will not function. If pressure 

monitoring is needed then change to the backup Console, 

Motor and Flow Probe. Switch to the backup System 

according the procedure described in Section 2. Resume 

support. Record the alert message and contact your Abbott 

Medical representative.

P4

Alert

PRESSURE 

1 BELOW 

MINIMUM

Check for and resolve a physiological or mechanical cause. 

Ensure appropriate pressure alarms are set. Consider re-

calibrating the transducer, if alert cannot be explained by 

conventional troubleshooting.

P5

Alert

PRESSURE 

2 BELOW 

MINIMUM

Check for and resolve a physiological or mechanical cause. 

Ensure appropriate pressure alarms are set. Consider re-

calibrating the transducer, if alert cannot be explained by 

conventional troubleshooting.

P6

Alert

PRESSURE 

1 ABOVE 

MAXIMUM

Check for and resolve a physiological or mechanical cause. 

Ensure appropriate pressure alarms are set. Consider re-

calibrating the transducer, if alert cannot be explained by 

conventional troubleshooting. Consider reducing RPM to 

reduce the pressure if appropriate.

P7

Alert

PRESSURE 

2 ABOVE 

MAXIMUM

Check for and resolve a physiological or mechanical cause. 

Ensure appropriate pressure alarms are set. Consider re-

calibrating the transducer, if alert cannot be explained by 

conventional troubleshooting. Consider reducing RPM to 

reduce the pressure if appropriate.

M6

Alert

MOTOR 

OVER TEMP

Switch to backup Console, Motor and Flow Probe accord-

ing to the procedure described in Section 2. Verify that the 

backup Motor stands free and is not covered (e.g. blankets).

B3

Alert

BATTERY 

CHARGER 

FAIL

Press the alarm acknowledge button. If the alert message 

repeats, switch to backup System as described in Section 

2. If this alert is associated with BATTERY MODULE FAIL 

then carry out the procedure associated with that alarm.

B4

Alert

BATTERY 

MAINTE-

NANCE 

REQUIRED

Do not use the Console. Perform battery maintenance ac-

cording to the Battery Maintenance Procedure listed in the 

CentriMag Circulatory Support System Operation Manual.

B5

Alert

LOW

BATTERY

Plug the Console into a AC power outlet to charge the 

battery. If no AC outlet is available, switch to the backup 

Console, Motor and Flow Probe according to the procedure 

described in Section 2.

Resume support.

B6

Alert

ON BAT-

TERY

Verify that the user wants the Console to be on battery. If so, 

carefully monitor the status of the battery charge indicator, 

while using the System on battery.

Reconnect to AC-outlet, as soon as possible.

CentriMag

TM 

Circulatory Support System

Alarms and Alerts Quick Reference Guide

 

RESPONSE TO SYSTEM ALARMS OR ALERTS

When a System alarm or alert condition occurs, an alarm or alert tone sounds and 

text message appears on the display of the 2nd Generation CentriMag Primary 

Console. Most alarms and alerts require some action on the user’s part to correct 

the cause. The following table can be used to determine how to respond to an 

alarm or alert condition. 