36
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1
What's New
What's New in Version 7.30
New features in version 7.30 of the InfiniiVision M9241/42/43A PXIe oscilloscope
software are:
• New Slew Rate and Y at X measurements along with improvements to the Time
at Edge measurement.
• The new Digitizer (acquisition) mode lets you choose the acquisition sample
rate and memory depth instead of having them be determined automatically
based on the horizontal time per division setting.
• Added support for getting waveform data for all segments at once (when in the
segmented acquisition mode).
• Added support for licenses that allow support subscriptions.
More detailed descriptions of the new and changed commands appear below.
New Commands
Changed
Commands
Command
Description
:ACQuire:DIGitizer (see
)
Turns Digitizer mode on or off.
:ACQuire:POINts[:ANALog]:AUT
O (see
)
Enables or disables Automatic determination of the analog
channel memory depth.
:ACQuire:SRATe[:ANALog]:AUT
O (see
)
Enables or disables Automatic determination of the analog
channel sampling rate.
:MEASure:SLEWrate (see
)
Installs a slew rate measurement on screen or returns the
measured value.
:MEASure:YATX (see
)
Installs a Y at X (vertical value at horizontal location)
measurement on screen or returns the measured value.
:SYSTem:DIDentifier? (see
Returns the oscilloscope's Host ID string.
:WAVeform:SEGMented:ALL
(see
)
Enables or disables the "waveform data for all segments" setting.
:WAVeform:SEGMented:XLISt?
(see
)
Returns the X (time) information for all segments at once.
Command
Differences
:ACQuire:POINts[:ANALog] (see
)
This command can now set a memory depth and turn on Digitizer
mode. Previously, this was a query only.
:ACQuire:SRATe[:ANALog] (see
)
This command can now set a sample rate and turn on Digitizer
mode. Previously, this was a query only.