:ACQuire:POINts[:ANALog]:AUTO
Command Syntax
:ACQuire:POINts[:ANALog]:AUTO {{0 | OFF} | {1 | ON}}
The :ACQuire:POINts[:ANALog]:AUTO command enables or disables Automatic
determination of the analog channel memory depth:
• ON — the analog channel memory depth is automatically determined by the
oscilloscope based on the horizontal time/div setting (Automatic mode, the
oscilloscope's default).
This is equivalent to turning Digitizer mode off using the :ACQuire:DIGitizer
command.
• OFF — the anallog channel memory depth is set using the
:ACQuire:POINts[:ANALog] command.
This is equivalent to turning Digitizer mode on using the :ACQuire:DIGitizer
command.
The Automatic or Digitizer mode setting also affects the sample rate (see the
:ACQuire:SRATe[:ANALog] command).
Query Syntax
:ACQuire:POINts[:ANALog]:AUTO?
The :ACQuire:POINts[:ANALog]:AUTO? query returns the automatic analog
memory depth setting.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also
•
•
•