:WGEN<w> Commands
:WGEN<w>:VOLTage:LOW
Command Syntax
:WGEN<w>:VOLTage:LOW <low>
<w> ::= 1 to (# WaveGen outputs) in NR1 format
<low> ::= low-level voltage in volts, in NR3 format
For all waveforms except DC, the :WGEN<w>:VOLTage:LOW command specifies
the waveform's low-level voltage. Use the :WGEN<w>:VOLTage:HIGH command
to specify the high-level voltage.
You can also specify the high-level and low-level voltages indirectly using the
:WGEN<w>:VOLTage and :WGEN<w>:VOLTage:OFFSet commands. For example, a
high-level voltage of 4 V and a low-level voltage of -1 V is the same as an
amplitude of 5 V and an offset of 1 V.
Query Syntax
:WGEN<w>:VOLTage:LOW?
The :WGEN<w>:VOLTage:LOW? query returns the currently specified waveform
low-level voltage.
Return Format
<low><NL>
<low> ::= low-level voltage in volts, in NR3 format
See Also
•
"Introduction to :WGEN<w> Commands"
•
•
•
•
NOTE
When the amplitude is below 40 mV, the offset is limited to ±500 mV.