:TIMebase Commands
30
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1059
:TIMebase:WINDow:POSition
Command Syntax
:TIMebase:WINDow:POSition <pos value>
<pos value> ::= time from the trigger event to the zoomed (delayed)
view reference point in NR3 format
The :TIMebase:WINDow:POSition command sets the horizontal position in the
zoomed (delayed) view of the main sweep. The main sweep range and the main
sweep horizontal position determine the range for this command. The value for
this command must keep the zoomed view window within the main sweep range.
Query Syntax
:TIMebase:WINDow:POSition?
The :TIMebase:WINDow:POSition? query returns the current horizontal window
position setting in the zoomed view.
Return Format
<value><NL>
<value> ::= position value in seconds
See Also
•
"Introduction to :TIMebase Commands"
•
•
•
•
•
•