:MEASure Commands
:MEASure:XMAX
Command Syntax
:MEASure:XMAX [<source>]
<source> ::= {CHANnel<n> | FUNCtion<m> | MATH<m> | FFT | WMEMory<r>}
<n> ::= 1-2 or 1-4 (# of analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:XMAX command installs a screen measurement and starts an
X-at-Max-Y measurement on the selected window. If the optional source
parameter is specified, the current source is modified.
Query Syntax
:MEASure:XMAX? [<source>]
The :MEASure:XMAX? query measures and returns the horizontal axis value at
which the maximum vertical value occurs. If the optional source is specified, the
current source is modified.
Return Format
<value><NL>
<value> ::= horizontal value of the maximum in NR3 format
See Also
•
"Introduction to :MEASure Commands"
•
•
NOTE
:MEASure:XMAX is an alias for :MEASure:TMAX.