Monitor Operating Keys
B
Alarm Silence
Use this key to silence the alarms and to cancel the alarm silence.
Menu / Enter Key
Press this key to enter the menu system, and to advance to the next
menu selection.
Cancel / Exit Key
Press this key to exit the menu system.
n
Up and
o
Down Arrows
Use the up and down arrow keys to adjust pulse beep volume during
normal operation. In a menu, use these keys to adjust the selection.
x
ON/OFF Key
Momentarily press this key when the device is OFF to turn the monitor ON.
Press and hold this key when the device is OFF to clear the patient trend data.
Silencing Alarms
Audible alarms can be silenced for two (2) minutes.
1.
To silence the alarm tone, press the
2.
Alarm Silence
key (
B
).
The Alarm Silence indicator will light solid for the first minute and 45 seconds,
3.
then flash the last 15 seconds.
To cancel the 2-minute silence, press the
4.
Alarm Silence
key (
B
).
NOtE: the Alarm Silence key will not affect visual alarm indicators or pulse tone.
NOtE: Permanent Alarm Silence is not available.
Changing the Alarm Limits
To select alarm limit defaults by patient type, press the
1.
Menu
key (
) once to display the current patient type
setting.
Pt
will be displayed as well as the current setting: Horse (
KrS
), Dog (
d09
), Cat (
[At
), Factory (
FA[
), or
Custom (
[Us
).
Use the
2.
Up
and
Down
keys (
n
o
)to select the desired patient type.
Press the
3.
Menu
key (
) again to display the desired alarm limit values for that patient type.
Use the
4.
Up
and
Down
keys (
n
o
) to increase or decrease the setting.
Press the
5.
Menu
key (
) to advance to the next alarm limit.
To exit, press the
6.
Exit
key (
) or wait 20 seconds until the patient reading is displayed.
Page 1 of 2
6
For use by or on the order of a licensed veterinarian.
f
Attention, see operation manual for additional instructions for use.
Quick Reference Guide