Pulse Volume
Use
1.
Up
and
Down
keys (
n
o
) to raise or lower pulse beep volume.
Continue to press the
2.
Down
key to silence the pulse tone completely.
Entering Setup Menu
Power oximeter on by pressing the
1.
On/Off
key (
x
) located on the device’s upper right side.
When the patient number (Example:
2.
P1
) is displayed, press and hold the
Exit
key (
) and then
press the
Menu
key (
).
The Averaging Menu will be the first option.
3.
AVG
will be displayed. Release the
Exit
key (
).
Setting Averaging
While
1.
AVG
is being displayed, use the
Up
and
Down
keys (
n
o
) to select the desired SpO
2
averaging (1-4).
Press the
2.
Menu
key (
) to advance to the next menu.
To exit menus, press the
3.
Exit
key (
) or wait 20 seconds and menu will be exited automatically.
Setting Sensitivity
Advance from previous menu or see
1.
Entering Setup Menu
.
KS
(for High Sensitivity) will be displayed.
Use the
2.
Up
and
Down
keys (
n
o
) to turn the High Sensitivity setting
0n
or
0ff
.
Display Brightness
Advance from previous menu or see
1.
Entering Setup Menu
. The display will show
db
.
Use the
2.
Up
and
Down
keys (
n
o
) to select the desired brightness, ranging from 1-10. The default setting is 10
(brightest).
Setting the Date
Advance from previous menu or see
1.
Entering Setup Menu
. To skip initial menus, if setting only date, press
Menu
key
(
) three (3) times until
dt1
(Date Setting 1) is displayed.
Use the
2.
Up
and
Down
keys (
n
o
) to select the year. Range is 00 (2000) to 99 (2099).
Press the
3.
Menu
key (
) to advance to the Month Menu. The display will show
dt2
(Date Setting 2).
Use the
4.
Up
and
Down
keys (
n
o
) to select the month. Range is 01 (January) to 12 (December).
Press the
5.
Menu
key (
) to advance to the Day Menu. The display will show
dt3
(Date Setting 3).
Use the
6.
Up
and
Down
keys (
n
o
) to select the day. Range is from 01 to 31.
Setting the time
Advance from previous menu or see
1.
Entering Setup Menu
. To skip initial menus, if setting only time, press
Menu
key
(
) six (6) times until
t1
(Time Setting 1) is displayed.
The display will show
2.
t1
. Use the
Up
and
Down
keys (
n
o
) to select the hour. Range is 00 (Midnight)
to 23 (11PM).
Press the
3.
Menu
key (
) to advance to Minute Menu. The display will show
t2
. Range is 0-59.
Press
4.
Exit
key (
) to exit Setup Menus.
f
Attention, see operation manual for Warnings, Cautions, Notes and additional instructions for use.
Page 2 of 2
Quick Reference Guide
V1030
Manufactured By: Smiths Medical PM, Inc.
N7W22025 Johnson Drive
Waukesha, Wisconsin 53186-1856 USA
Phone: 1-262-513-8500
Toll-Free: (U.S.A. only) 1-888-745-6562
Fax: 1-262-542-0718
www.surgivet.com
SurgiVet and the Smiths design mark are trademarks of the Smiths Medical family of companies. The
symbol
®
indicates the trademark is registered in the U.S. Patent and Trademark Office and certain
other countries.
©2010 Smiths Medical Family of companies. All rights reserved.
H