1xEV-DO Signaling
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
221
User Manual 1173.9511.02 ─ 11
Manual operation:
See
2.6.11.3
Measurement States
The following commands return the current measurement states.
.............................................................................. 221
FETCh:EVDO:SIGN<i>:THRoughput:STATe?
...................................................................221
FETCh:EVDO:SIGN<i>:PER:STATe:ALL?
........................................................................221
FETCh:EVDO:SIGN<i>:THRoughput:STATe:ALL?
............................................................221
FETCh:EVDO:SIGN<i>:PER:STATe?
FETCh:EVDO:SIGN<i>:THRoughput:STATe?
Queries the main measurement state. Use
FETCh:...:STATe:ALL?
to query the
measurement state including the substates. Use
INITiate..., STOP...,
ABORt...
to change the measurement state.
See also: "Measurement Control" in the R&S
CMW user manual, chapter "Remote
Control"
Return values:
<State>
OFF | RDY | RUN
OFF:
measurement switched off, no resources allocated, no
results available (when entered after
ABORt...
)
RDY:
measurement has been terminated, valid results may be
available
RUN:
measurement running (after
INITiate..., READ...
),
synchronization pending or adjusted, resources active or queued
Example:
See
Perform Forward Link Measurements
Usage:
Query only
Firmware/Software:
V2.1.20
Options:
R&S CMW-KS890 required for Throughput measurements
Manual operation:
See
FETCh:EVDO:SIGN<i>:PER:STATe:ALL?
FETCh:EVDO:SIGN<i>:THRoughput:STATe:ALL?
Queries the main measurement state and the measurement substates. Both measure-
ment substates are relevant for running measurements only. Use
FETCh:...:STATe?
to query the main measurement state only. Use
INITiate...,
STOP..., ABORt...
to change the measurement state.
See also: "Measurement Control" in the R&S
CMW user manual, chapter "Remote
Control"
Command Reference