CDMA2000 Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
790
User Manual 1173.9511.02 ─ 11
..............................................................................................819
...........................................................................................821
........................................................................................... 824
................................................................................................ 826
........................................................................................... 826
.......................................................................................... 831
Modulation Results (Single Values)
...................................................................... 831
................................................................................ 833
.............................................................................. 835
List Mode Results: Result Groups, Single Segment
............................................. 836
List Mode Results: Result Groups, All Segments
................................................. 852
List Mode Results: Single Results, All Segments
................................................. 867
5.5.4.1
Measurement Control and States
The following commands control the measurement and return the current measure-
ment state.
INITiate:CDMA:MEAS<i>:MEValuation
.............................................................................790
............................................................................... 790
ABORt:CDMA:MEAS<i>:MEValuation
..............................................................................790
FETCh:CDMA:MEAS<i>:MEValuation:STATe?
.................................................................791
FETCh:CDMA:MEAS<i>:MEValuation:STATe:ALL?
.......................................................... 791
INITiate:CDMA:MEAS<i>:MEValuation
STOP:CDMA:MEAS<i>:MEValuation
ABORt:CDMA:MEAS<i>:MEValuation
Starts, stops, or aborts the measurement:
●
INITiate...
starts or restarts the measurement; the R&S
CMW enters the
"RUN" state.
●
STOP...
causes a running measurement to stop after the current evaluation
period is terminated and valid results are available; the R&S
CMW enters the
"RDY" state.
●
ABORt...
causes a running measurement to stop immediately; the R&S
CMW
enters the "OFF" state.
Use
FETCh...STATe?
to query the current measurement state.
See also: "Measurement Control" in the R&S
CMW user manual, chapter "Remote
Control"
Example:
See
Multi Evaluation: Single-Shot and Continuous Measure-
Usage:
Event
Firmware/Software:
V1.0.2.7
Manual operation:
See
Command Reference