1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
442
User Manual 1173.9511.02 ─ 11
If the first measurement cycle has not been completed when the timer expires, the
measurement is stopped. The measurement state changes to
RDY
and the reliability
indicator is set to 1, indicating that a measurement timeout occurred. Still running
READ
,
FETCh
or
CALCulate
commands are completed, returning the available results.
At least for some results there are no values at all or the statistical depth has not been
reached.
A timeout of 0 s corresponds to an infinite measurement timeout.
Parameters:
<Timeout>
Default unit: s
Example:
See
OLTR: Specifying Measurement-Specific Settings
Firmware/Software:
V3.0.10
Manual operation:
See
CONFigure:EVDO:MEAS<i>:OLTR:REPetition?
Specifies the repetition mode of the measurement. The repetition mode specifies
whether the measurement is stopped after a single-shot or repeated continuously. Use
CONFigure:..:MEAS<i>:...:SCOunt
to determine the number of measurement
intervals per single shot.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
Parameters:
<Repetition>
SINGleshot | CONTinuous
SINGleshot:
Single-shot measurement
CONTinuous:
Continuous measurement
*RST:
SING
Example:
See
OLTR: Specifying Measurement-Specific Settings
Usage:
Query only
Firmware/Software:
V3.0.10
CONFigure:EVDO:MEAS<i>:OLTR:SEQuence
<NoOfMeasSeq>
Sets/gets the number of measurement sequences within a single OLTR measurement.
Each sequence consists of a power UP or power DOWN step, followed by a power
step in the opposite direction (see
Parameters:
<NoOfMeasSeq>
Range:
1 to 5
*RST:
1
Example:
See
OLTR: Specifying Measurement-Specific Settings
Firmware/Software:
V3.0.10