1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
377
User Manual 1173.9511.02 ─ 11
READ:EVDO:MEAS<i>:MEValuation:TRACe:ACP:EXTended:AVERage:
ABSolute?
READ:EVDO:MEAS<i>:MEValuation:TRACe:ACP:EXTended:MAXimum:
ABSolute?
Returns the adjacent channel absolute power measured in dBm at a series of frequen-
cies. The current, average and maximum traces can be retrieved.
The frequencies are determined by the offset values defined via the commands
and
. All defined
offset values are considered (irrespective of their activation status).
The values described below are returned by
FETCh
and
READ
commands.
CALCulate
commands return limit check results instead, one value for each result listed below.
Return values:
<1_Reliability>
<2_PowOffsetM19>
...
<21_PowOffsetM0>
<23_PowCarrier>
<24_PowOffsetP0> ...
<43_PowOffsetP19>
The power offsets <PowOffsetPn> (<PowOffsetMn>) refer to the
average absolute power at the carrier frequency plus (minus) the
frequency offset value number n.
Range:
-100 dBm to 20 dBm
Default unit: dBm
Example:
See
Multi Evaluation: Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V3.2.70
For additional information concerning syntax elements and returned values refer to
Conventions and General Information
.
CALCulate:EVDO:MEAS<i>:MEValuation:TRACe:ACP:EXTended:AVERage[:
RELative]?
CALCulate:EVDO:MEAS<i>:MEValuation:TRACe:ACP:EXTended:CURRent[:
RELative]?
CALCulate:EVDO:MEAS<i>:MEValuation:TRACe:ACP:EXTended:MAXimum[:
RELative]?
FETCh:EVDO:MEAS<i>:MEValuation:TRACe:ACP:EXTended:CURRent[:
RELative]?
FETCh:EVDO:MEAS<i>:MEValuation:TRACe:ACP:EXTended:AVERage[:
RELative]?
FETCh:EVDO:MEAS<i>:MEValuation:TRACe:ACP:EXTended:MAXimum[:
RELative]?
READ:EVDO:MEAS<i>:MEValuation:TRACe:ACP:EXTended:CURRent[:
RELative]?
Command Reference