1xEV-DO Signaling
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
224
User Manual 1173.9511.02 ─ 11
Firmware/Software:
V2.1.20
Options:
R&S CMW-KS890
Manual operation:
See
CONFigure:EVDO:SIGN<i>:RXQuality:FLPer:SCONdition
<StopCondition>
CONFigure:EVDO:SIGN<i>:RXQuality:RLPer:SCONdition
<StopCondition>
Qualifies whether the measurement is stopped after a failed limit check or continued.
SLFail
means that the measurement is stopped and reaches the
RDY
state as soon
as one of the results exceeds the limits.
Parameters:
<StopCondition>
NONE | ALEXceeded | MCLexceeded | MPERexceeded
NONE
: Continue measurement irrespective of the limit check
ALEXceeded
: Stop if any limit is exceeded
MCLexceeded
: Stop if minimum confidence level is exceeded
MPERexceeded
: Stop if max. PER is exceeded
*RST:
NONE
Example:
See
Firmware/Software:
V2.1.20
Manual operation:
See
"Forward/Reverse Link PER > Stop Condition"
CONFigure:EVDO:SIGN<i>:RXQuality:FLPer:MTPSent
<MaxTestPackSent>
Defines the length of a single shot forward link PER measurement, i.e. the maximum
number of test packets sent.
Parameters:
<MaxTestPackSent> Range:
1 to 10000
*RST:
1000
Example:
See
Firmware/Software:
V1.0.15.0
Manual operation:
See
"Forward/Reverse Link PER > Max Test Packets Sent"
CONFigure:EVDO:SIGN<i>:RXQuality:RLPer:MPSent
<MaxPacketsSent>
Defines the length of a single shot reverse link PER measurement, i.e. the maximum
number of test packets sent by the AT.
Parameters:
<MaxPacketsSent>
Range:
1 to 10000
*RST:
1000
Example:
See
Firmware/Software:
V2.1.20