1xEV-DO Signaling
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
101
User Manual 1173.9511.02 ─ 11
................................................................................................ 102
................................................................................................102
............................................................................................102
Access Probes
Specifies whether the R&S
CMW acknowledges or ignores access probes from the AT.
●
Acknowledge
: Normal operation, needed for successful session negotiation and
connection setup.
●
Ignore
: The R&S
CMW does not acknowledge the access probes, forcing the AT
to send repeated access probes. The access probe power is exclusively controlled
by the open loop rule; hence it possible to test open loop power control of the AT
using an additional power meter.
Remote command:
Initial Adjust
Initial power offset for access probes (ProbeInitialAdjust parameter in the AccessPara-
meters Message). "Initial Adjust" is a correction factor to be used by access terminals
in the open loop power estimate, expressed as a two’s complement value in units of 1
dB.
Setting this parameter requires option R&S CMW-KS890.
Remote command:
Open Loop Adjust
Nominal transmit power offset (OpenLoopAdjust) to be used by ATs for the given band
class in the open loop power estimate, in steps of 1 dB.
Remote command:
Probe Increment
Step size of power increases (PowerStep) between consecutive access probes with a
resolution of 0.5 dB.
Setting this parameter requires option R&S CMW-KS890.
Remote command:
Probes per Sequence
Maximum number of access probes (ProbeNumStep) which ATs are to transmit in a
single access probe sequence.
Setting this parameter requires option R&S CMW-KS890.
Remote command:
GUI Reference