1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
326
User Manual 1173.9511.02 ─ 11
TRIGger:EVDO:MEAS<i>:MEValuation:THReshold
<Threshold>
Defines the trigger threshold for power trigger sources.
Parameters:
<Threshold>
Range:
-50 dB to 0 dB
*RST:
-30 dB
Default unit: dB
Example:
See
Configuring the Trigger System
Firmware/Software:
V2.0.10
Manual operation:
See
TRIGger:EVDO:MEAS<i>:MEValuation:DELay
<Delay>
Defines a time delaying the start of the measurement relative to the trigger event. This
setting has no influence on "Free Run" measurements.
Parameters:
<Delay>
Range:
-1.25E-3 s to 0.08 s
*RST:
0 s
Default unit: s
Example:
See
Configuring the Trigger System
Firmware/Software:
V2.0.20
Manual operation:
See
TRIGger:EVDO:MEAS<i>:MEValuation:TOUT
<TriggerTime>
Selects the maximum time that the R&S
CMW will wait for a trigger event before it
stops the measurement in remote control mode or indicates a trigger timeout in manual
operation mode. This setting has no influence on "Free Run" measurements.
Parameters:
<TriggerTime>
Range:
0 s to 83.3 s
*RST:
1 s
Default unit: s
Additional parameters: OFF | ON (disables | enables the time-
out)
Example:
See
Configuring the Trigger System
Firmware/Software:
V1.0.4.11
V2.1.20 Default unit changed from ms to s.
Manual operation:
See
Command Reference