1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
249
User Manual 1173.9511.02 ─ 11
Fig. 3-4: Example: Sequence with four segments and accuracy and not accuracy measurement
results.
The list mode is essentially a single-shot remote control application. When a measure-
ment is initiated in list mode, all defined segments are measured once. Afterwards the
results can be retrieved using FETCH COMMANDS. The parameters in the figures are
set by means of the following remote control commands:
Parameters
SCPI commands
Number of segments
Slots per segment,
power
and
frequency
(step
length)
Statistical length
Activate / deactivate
list
mode
Retrieve results
FETCh:EVDO:MEAS<i>:MEValuation:LIST:SEGMent<no>:...
FETCh:EVDO:MEAS<i>:MEValuation:LIST:...
See
List Mode Results: Result Groups, Single Segment
Results: Result Groups, All Segments
The list mode can be activated/deactivated via command (see table above).
Global and list mode parameters
The RF settings (expected power, RF frequency) and most of the "Measurement Con-
trol" settings (step length, steps per segment, averaging lengths, enable/disable
results) are special list mode settings. The R&S
CMW ignores the corresponding "multi
evaluation" parameters. All other settings are taken from the multi evaluation measure-
ment, e.g.:
●
Measure on Exception
●
Some "Modulation" and "Spectrum" settings
●
All limit settings
General Description