CDMA2000 Signaling
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
527
User Manual 1173.9511.02 ─ 11
Select as fixed Target
Sets the selected measurement as fixed target of the softkey. The softkey label indi-
cates the measurement name and switches directly to the selected target without
opening the dialog box.
You can still change the target measurement or re-enable the dialog box using the
configuration menu, see
chapter 4.3.11, "Shortcut Configuration"
Go to / Cancel
Click "Go to" to switch to the selected measurement or "Cancel" to abort.
4.3.4 General Settings
These settings are located at the very top of the configuration dialog.
Fig. 4-28: General settings
................................................................................................ 528
Scenario
In general, this parameter switches between test scenarios that require different sets of
parameters. Some parameters are only available for certain scenarios or have a differ-
ent meaning depending on the selected scenario.
The parameter is only configurable while the forward signal is switched off.
●
Standard Cell
: standard CDMA2000 cell with only one RF output path.
●
Hybrid Mode
●
Standard Cell Fading
:
"Standard Cell" plus fading and/or AWGN insertion.
Additional parameter "Fading" selects either external fading via a connected R&S
AMU200A or internal fading via an internal fader I/Q board.
Options R&S CMW-KE100 and R&S CMW-KE800 are required for internal fading.
For further reference:
●
●
chapter 4.2.5, "External Fading"
●
chapter 4.2.6, "Internal Fading"
Remote command:
GUI Reference