1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
441
User Manual 1173.9511.02 ─ 11
<SyncState>
PEND | ADJ | INV
PEND:
waiting for resource allocation, adjustment, hardware
switching ("pending")
ADJ:
all necessary adjustments finished, measurement running
("adjusted")
INV:
not applicable because <MainState>: OFF or RDY ("inva-
lid")
<ResourceState>
QUE | ACT | INV
QUE:
measurement without resources, no results available
("queued")
ACT:
resources allocated, acquisition of results in progress but
not complete ("active")
INV:
not applicable because <MainState>: OFF or RDY ("inva-
lid")
Example:
See
Usage:
Query only
Firmware/Software:
V3.0.10
Manual operation:
See
3.5.5.2
Measurement Parameters
The following commands define general settings for the multi evaluation measurement.
CONFigure:EVDO:MEAS<i>:OLTR:TOUT
........................................................................ 441
CONFigure:EVDO:MEAS<i>:OLTR:REPetition?
................................................................442
CONFigure:EVDO:MEAS<i>:OLTR:SEQuence
................................................................. 442
CONFigure:EVDO:MEAS<i>:OLTR:PSTep
.......................................................................443
CONFigure:EVDO:MEAS<i>:OLTR:PSTep:DIRection
........................................................443
CONFigure:EVDO:MEAS<i>:OLTR:RPINterval
................................................................. 443
CONFigure:EVDO:MEAS<i>:OLTR:RPINterval:TIME?
.......................................................444
CONFigure:EVDO:MEAS<i>:OLTR:GINTerval
.................................................................. 444
CONFigure:EVDO:MEAS<i>:OLTR:GINTerval:TIME?
........................................................444
CONFigure:EVDO:MEAS<i>:OLTR:MOEXception
.............................................................445
CONFigure:EVDO:MEAS<i>:OLTR:LIMit:ILOWer
..............................................................445
CONFigure:EVDO:MEAS<i>:OLTR:TOUT
<Timeout>
Defines a timeout for the measurement. The timer is started when the measurement is
initiated via a
READ
or
INIT
command. It is not started if the measurement is initiated
manually (ON/OFF key or RESTART/STOP key).
When the measurement has completed the first measurement cycle (first single shot),
the statistical depth is reached and the timer is reset.
Command Reference
135-1095-0799