CDMA2000 Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
767
User Manual 1173.9511.02 ─ 11
CONFigure:CDMA:MEAS:MEValuation:LIMit:ACP:RELative 5,0,-5,-10,-15,-20
CONFigure:CDMA:MEAS:MEValuation:LIMit:ACP:ABSolute 5,0,-5,-10, OFF, OFF
WAITKEY >Limit settings complete, press "OK" to start single-shot measurement<
// *************************************************************************
// Define OBW limit 2.96 MHz
// *************************************************************************
CONFigure:CDMA:MEAS:MEValuation:LIMit:OBW 2.96
5.4.1.5
Multi Evaluation: Performing Single-Shot Measurements
// *************************************************************************
// Enable all measurements
// *************************************************************************
CONFigure:CDMA:MEAS:MEValuation:RESult:EVMagnitude ON
CONFigure:CDMA:MEAS:MEValuation:RESult:MERRor ON
CONFigure:CDMA:MEAS:MEValuation:RESult:PERRor ON
CONFigure:CDMA:MEAS:MEValuation:RESult:IQ ON
CONFigure:CDMA:MEAS:MEValuation:RESult:ACP ON
CONFigure:CDMA:MEAS:MEValuation:RESult:CDP ON
CONFigure:CDMA:MEAS:MEValuation:RESult:CDE ON
CONFigure:CDMA:MEAS:MEValuation:RESult:CP ON
CONFigure:CDMA:MEAS:MEValuation:RESult:CPO ON
CONFigure:CDMA:MEAS:MEValuation:RESult:CTO ON
CONFigure:CDMA:MEAS:MEValuation:RESult:POWer ON
CONFigure:CDMA:MEAS:MEValuation:RESult:MODQuality ON
// *************************************************************************
// Start single-shot measurement, return magnitude error trace.
// Query the measurement state (should be "RDY").
// *************************************************************************
READ:CDMA:MEAS:MEValuation:TRACe:MERRor:AVERage?
FETCh:CDMA:MEAS:MEValuation:STATe?
// *************************************************************************
// Read the current traces obtained in the last measurement
// without re-starting the measurement
// *************************************************************************
FETCh:CDMA:MEAS:MEValuation:TRACe:CDP:ISIGnal:CURRent?
FETCh:CDMA:MEAS:MEValuation:TRACe:CDP:QSIGnal:CURRent?
FETCh:CDMA:MEAS:MEValuation:TRACe:CDE:ISIGnal:CURRent?
FETCh:CDMA:MEAS:MEValuation:TRACe:CDE:QSIGnal:CURRent?
FETCh:CDMA:MEAS:MEValuation:TRACe:CP:ISIGnal:CURRent?
FETCh:CDMA:MEAS:MEValuation:TRACe:CP:QSIGnal:CURRent?
FETCh:CDMA:MEAS:MEValuation:TRACe:CPO:ISIGnal:CURRent?
FETCh:CDMA:MEAS:MEValuation:TRACe:CPO:QSIGnal:CURRent?
FETCh:CDMA:MEAS:MEValuation:TRACe:CTO:ISIGnal:CURRent?
FETCh:CDMA:MEAS:MEValuation:TRACe:CTO:QSIGnal:CURRent?
Programming