1xEV-DO Signaling
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
161
User Manual 1173.9511.02 ─ 11
Parameters:
<InsertLossMode>
NORMal | USER
NORMal
: the insertion loss is determined by the fading profile
USER
: the insertion loss can be adjusted by the user
*RST:
NORM
Example:
See
Firmware/Software:
V3.2.80
Options:
R&S CMW-KE100 and R&S CMW-KE800
Manual operation:
See
CONFigure:EVDO:SIGN<i>:FADing:FSIMulator:RESTart
Restarts the fading process in
MANual
mode (see
).
Usage:
Event
Firmware/Software:
V3.2.80
Options:
R&S CMW-KE100 and R&S CMW-KE800
Manual operation:
See
CONFigure:EVDO:SIGN<i>:FADing:FSIMulator:RESTart:MODE
<RestartMode>
Sets the restart mode of the fading simulator.
Parameters:
<RestartMode>
AUTO | MANual
AUTO
: fading automatically starts with the DL signal
MANual
: fading is started and restarted manually (see
*RST:
AUTO
Example:
See
Firmware/Software:
V3.2.80
Options:
R&S CMW-KE100 and R&S CMW-KE800
Manual operation:
See
CONFigure:EVDO:SIGN<i>:FADing:FSIMulator:STANdard
<Standard>
Selects one of the propagation condition defined in the table 6.4.1-1 of 3GPP2
C.S0032.
Command Reference
深圳德标仪器
135-1095-0799